SMALL BATTERY DRIVE II
Report
- Report Number
- 2520274-2014-11588
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- May 19, 2014
- Report Date
- May 19, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HWE
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
THE MANUFACTURING LOCATION WAS UNKNOWN. DEVICE MANUFACTURE DATE: THE DEVICE MANUFACTURE DATE IS CURRENTLY UNAVAILABLE. THIS DEVICE WAS RETURNED FOR SERVICE, HOWEVER DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMMERSION DURING CLEANING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE MANUFACTURER LOCATION WAS DOCUMENTED AS UNKNOWN IN THE INITIAL REPORT. THE LOCATION HAS BEEN UPDATED TO (B)(4). CONTACT OFFICE NAME/ADDRESS HAS BEEN UPDATED ACCORDINGLY TO REFLECT THE CORRECT MANUFACTURING FACILITY. DEVICE MANUFACTURE DATE: THE DEVICE MANUFACTURE DATE WAS DOCUMENTED AS UNKNOWN IN THE INITIAL REPORT. THE DEVICE MANUFACTURE DATE HAS BEEN UPDATED AS FEB 11, 2013. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS DISCOVERED THAT THE TRIGGERS ON THE SMALL BATTERY DRIVE DEVICE WERE STICKY. THE EVENT WAS NOT RELATED TO SURGERY. THERE WERE NO INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329134 | SMALL BATTERY DRIVE II | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT | HWE | DEPUY SYNTHES POWER TOOLS | 1078 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |