M2A 38MM MOD HD STD NK
Report
- Report Number
- 0001825034-2014-05267
- Event Type
- Injury
- Date Received
- June 5, 2014
- Date of Event
- August 26, 2014
- Report Date
- May 28, 2015
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- LZO
- PMA / PMN Number
- PK062997
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- ATTORNEY
Narratives
THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY CORRECTED INFORMATION.
THE FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION THAT WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: "ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL ON METAL ARTICULATING SURFACES." THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-05267 & 02515)
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 6 STATES, "INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS." NUMBER 14 STATES, "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN." THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
LEGAL COUNSEL REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2006. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTS PATIENT ALLEGATIONS OF PAIN, SWELLING, INFLAMMATION, AND LACK OF MOBILITY. THERE HAS BEEN NO REPORTED REVISION PROCEDURE TO DATE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. ADDITIONAL INFORMATION RECEIVED IN THE PATIENT'S REVISION OPERATIVE REPORT NOTED PATIENT UNDERWENT AN INITIAL LEFT HIP ARTHROPLASTY ON (B)(6) 2007. SUBSEQUENTLY, PATIENT UNDERWENT A BILATERAL HIP REVISION ON (B)(6) 2014. OPERATIVE REPORT FURTHER NOTED THE REASON FOR THE REVISION WAS HIGH CHROMIUM LEVELS. DURING THE REVISION, GRAYISH TYPE FLUID WAS FOUND IN THE LEFT HIP. THE MODULAR HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED. ON THE RIGHT HIP, THE MODULAR HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED AS WELL.
LEGAL COUNSEL REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2006. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTS PATIENT ALLEGATIONS OF PAIN, SWELLING, INFLAMMATION, AND LACK OF MOBILITY. THERE HAS BEEN NO REPORTED REVISION PROCEDURE TO DATE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. ADDITIONAL INFORMATION RECEIVED IN THE PATIENT'S REVISION OPERATIVE REPORT NOTED PATIENT UNDERWENT AN INITIAL RIGHT HIP ARTHROPLASTY ON (B)(6) 2007. SUBSEQUENTLY, PATIENT UNDERWENT A BILATERAL HIP REVISION ON (B)(6) 2014. OPERATIVE REPORT FURTHER NOTED THE REASON FOR THE REVISION WAS HIGH CHROMIUM LEVELS. DURING THE REVISION, GRAYISH TYPE FLUID WAS FOUND IN THE LEFT HIP. THE MODULAR HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED. ON THE RIGHT HIP, THE MODULAR HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED AS WELL.
LEGAL COUNSEL REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2006. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTS PATIENT ALLEGATIONS OF PAIN, SWELLING, INFLAMMATION, AND LACK OF MOBILITY. THERE HAS BEEN NO REPORTED REVISION PROCEDURE TO DATE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329207 | M2A 38MM MOD HD STD NK | PROSTHESIS, HIP | LZO | BIOMET ORTHOPEDICS | N/A | 628970 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Required Intervention |