FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3851984 · Received June 5, 2014

Report

Report Number
1416980-2014-18052
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 13, 2014
Report Date
May 13, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. A VISUAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. A CLEAR PASSAGE TEST AND PRESSURE TEST WERE PERFORMED WITH NO ISSUES NOTED. A DIMENSIONAL ANALYSIS WAS PERFORMED ON THE SPIKE PORT WITH ALL MEASUREMENTS WITHIN SPECIFICATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS REPORTED TO BE AVAILABLE BUT HAS NOT YET BEEN RECEIVED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LEAK WAS OBSERVED BETWEEN THE CONNECTION OF THE CASSETTE AND A SOLUTION BAG. THE EVENT OCCURRED DURING PRIME ON THE HOME CHOICE (HC). NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329120 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE