HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-18052
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 13, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. A VISUAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. A CLEAR PASSAGE TEST AND PRESSURE TEST WERE PERFORMED WITH NO ISSUES NOTED. A DIMENSIONAL ANALYSIS WAS PERFORMED ON THE SPIKE PORT WITH ALL MEASUREMENTS WITHIN SPECIFICATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). THE DEVICE WAS REPORTED TO BE AVAILABLE BUT HAS NOT YET BEEN RECEIVED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A LEAK WAS OBSERVED BETWEEN THE CONNECTION OF THE CASSETTE AND A SOLUTION BAG. THE EVENT OCCURRED DURING PRIME ON THE HOME CHOICE (HC). NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329120 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE |