FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3851941 · Received June 5, 2014

Report

Report Number
1416980-2014-18045
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
March 23, 2014
Report Date
May 12, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED FOR THE REPORTED EVENT. A REVIEW OF THE EVENT HISTORY LOG VERIFIED THE OCCURRENCE OF AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE THAT ONE OR MORE CYCLES ADVANCED TO THE NEXT FILL WHEN SLOW/NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 00:53:13. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1248ML, INDICATING THE HOME PATIENT (HP) DRAINED 1248ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328538 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1