FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3851787 · Received April 11, 2014

Report

Report Number
8020893-2014-00888
Event Type
Malfunction
Date Received
April 11, 2014
Date of Event
March 13, 2014
Report Date
March 14, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB). THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. COVIDIEN REF: (B)(4).

Description of Event or Problem · 1

COVIDIEN RECEIVED INFORMATION STATING THAT AN 840 VENTILATOR EXPERIENCED DEVICE ALERTS RENDERING THE VENTILATOR INOPERABLE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
221215 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1