FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3851780
·
Received April 11, 2014
Report
- Report Number
- 8020893-2014-00873
- Event Type
- Malfunction
- Date Received
- April 11, 2014
- Date of Event
- March 6, 2014
- Report Date
- March 13, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB). THE CSE UPGRADED THE SOFTWARE TO THE CURRENT REVISION AND PERFORMED THE EXTENDED SELF-TEST (EST), SHORT SELF-TEST (SST) AND THE PERFORMANCE VERIFICATION TEST (PVT); THE DEVICE OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS. COVIDIEN REF: (B)(4).
Description of Event or Problem · 1
A REPORT RECEIVED FROM THE USA STATED AN 840 VENTILATOR EXPERIENCED A DEVICE ALERT AND SUBSEQUENTLY STOPPED CYCLING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 221216 | 840 VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |