FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3851780 · Received April 11, 2014

Report

Report Number
8020893-2014-00873
Event Type
Malfunction
Date Received
April 11, 2014
Date of Event
March 6, 2014
Report Date
March 13, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB). THE CSE UPGRADED THE SOFTWARE TO THE CURRENT REVISION AND PERFORMED THE EXTENDED SELF-TEST (EST), SHORT SELF-TEST (SST) AND THE PERFORMANCE VERIFICATION TEST (PVT); THE DEVICE OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS. COVIDIEN REF: (B)(4).

Description of Event or Problem · 1

A REPORT RECEIVED FROM THE USA STATED AN 840 VENTILATOR EXPERIENCED A DEVICE ALERT AND SUBSEQUENTLY STOPPED CYCLING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
221216 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1