FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3851722 · Received June 5, 2014

Report

Report Number
1416980-2014-18022
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
January 13, 2014
Report Date
May 13, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE REPORTED IIPV CONDITION. THE CAUSE WAS USE ERROR, TIDAL TOTAL UF REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014, AT 01:49:42. DURING NIGHT DRAIN CYCLE TEN, THE PATIENT'S ULTRAFILTRATION READING WAS 1904ML, INDICATING THE HOME PATIENT (HP) DRAINED 1364ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1800ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328941 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1