HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-18016
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- February 10, 2014
- Report Date
- May 12, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4).THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE REPORTED IIPV CONDITION. THE CAUSE WAS DETERMINED TO BE A USE ERROR, TIDAL TOTAL UF REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 20:12:26. DURING NIGHT DRAIN CYCLE 17, THE PATIENT'S ULTRAFILTRATION READING WAS 173ML, INDICATING THE HOME PATIENT (HP) DRAINED 130ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 170ML. NO ADDITIONAL INFORMATION IS AVAILABLE.