FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3851431 · Received June 5, 2014

Report

Report Number
1416980-2014-17991
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
April 14, 2014
Report Date
May 12, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS COMPLETED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. THE CAUSE OF THE EVENT WAS A FALSE EMPTY DETECT AND USE ERROR, FURTHER SPECIFIED AS THE THE MINIMUM DRAIN VOLUME PERCENT WAS SET TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER'S GUIDE PROVIDES A WARNING "IF YOU SET THE MINIMUM DRAIN VOLUME % TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THERAPY INITIATED ON (B)(6) 2014 AT 22:22:29. DURING NIGHT DRAIN CYCLE SEVEN, THE PATIENT'S ULTRAFILTRATION READING WAS 1439ML, INDICATING THE HOME PATIENT (HP) DRAINED 1439ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1400ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328664 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1