LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-00858
- Event Type
- Malfunction
- Date Received
- March 26, 2014
- Date of Event
- February 18, 2014
- Report Date
- March 26, 2014
- Manufacturer
- ZOLL LIFECOR CORP.
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK THERAPY ELECTRODE MESSAGES, CONNECTOR WON'T LATCH) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST AND THE TRUNK CABLE CONNECTOR WOULDN'T LATCH. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE IN THE CABLE CONNECTING THE DISTRIBUTION NODE AND ECG "B". THERE WAS NO SIGN OF EXTERNAL DAMAGE TO THE CABLE. THE ROOT CAUSE FOR THE OPEN PULSE WIRE AND DEFECTIVE TRUNK CABLE CONNECTOR COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
THE WIFE OF A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT WAS RECEIVING CHECK THERAPY ELECTRODE ALARMS AND HIS ELECTRODE BELT WOULDN'T STAY CONNECTED TO HIS MONITOR. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 176635 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORP. | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |