FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 3851251 · Received March 25, 2014

Report

Report Number
9615739-2014-00132
Event Type
Malfunction
Date Received
March 25, 2014
Date of Event
February 25, 2014
Report Date
February 25, 2014
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND ONE BRAKE AND ONE STEER CASTER ON FOOT END OF BED INOPERATIVE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED ONE BRAKE AND ONE STEER CASTER ON FOOT END OF BED RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES ON FOOT END OF THE BED DO NOT HOLD. THE BED WAS LOCATED AT THE ACCOUNT IN THE 4TH FLOOR BED STORAGE. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174246 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1