FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3851091 · Received June 5, 2014

Report

Report Number
3004209178-2014-85221
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 15, 2014
Report Date
May 15, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNABLE TO PRIME THE INSULIN PUMP DURING THE PRIME TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. CRACKS TO THE CASE, RESERVOIR TUBE LIP, BELT CLIP SLOT, A SCRATCHED LCD WINDOW AND MISSING END CAP STICKER ALSO NOTED.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP WAS ALARMING DURING THE PRIMING PROCESS. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 187 MG/DL. THE CUSTOMER ALSO REPORTED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED. THE CUSTOMER REPORTED HAVING DROPPED THE INSULIN PUMP. THE INSULIN PUMP WILL BE REPLACED. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329625 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 46 YR