FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3851038 · Received February 12, 2014

Report

Report Number
2518422-2014-00189
Event Type
Malfunction
Date Received
February 12, 2014
Date of Event
January 20, 2014
Report Date
January 20, 2014
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S POWER MANAGEMENT BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93623 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054260

Patients

Seq Age Sex Outcome Treatment
1