FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUALCHAN INF

MDR report key: 3850926 · Received February 12, 2014

Report

Report Number
9615050-2014-01146
Event Type
Malfunction
Date Received
February 12, 2014
Date of Event
January 1, 2014
Report Date
January 14, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0070-20113
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE DID NOT PASS TOUCHSCREEN TEST. THIS WAS DUE TO RESIDUE ON THE TOUCHSCREEN FROM THE PRODUCT THE CUSTOMER USED TO CLEAN THE DEVICE. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN NEEDS CALIBRATING. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE TOUCHSCREEN TEST. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93778 SYMBIQ DUALCHAN INF 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA