FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ DUALCHAN INF
MDR report key: 3850926
·
Received February 12, 2014
Report
- Report Number
- 9615050-2014-01146
- Event Type
- Malfunction
- Date Received
- February 12, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 14, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0070-20113
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE DID NOT PASS TOUCHSCREEN TEST. THIS WAS DUE TO RESIDUE ON THE TOUCHSCREEN FROM THE PRODUCT THE CUSTOMER USED TO CLEAN THE DEVICE. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN NEEDS CALIBRATING. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE TOUCHSCREEN TEST. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93778 | SYMBIQ DUALCHAN INF | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |