FDA Adverse Event Malfunction Summary report: N

SYMBIQ 3.13 DUAL CHA

MDR report key: 3850907 · Received February 12, 2014

Report

Report Number
9615050-2014-01140
Event Type
Malfunction
Date Received
February 12, 2014
Date of Event
January 2, 2014
Report Date
January 14, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN WAS INOPERABLE AND THE DEVICE WAS OVERHEATING. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPT WITH AN UNSIGNED NOTE THAT STATED, "HIGH TEMP AND TOUCH INOP." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DOES NOT RESPOND WHEN PRESSED AND HAD AN UNSPECIFIED OVERHEATING CHARRING OR MELTING NOTED ON THE DEVICE AND THE DEVICE WAS NOT HOT TO TOUCH. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
92931 SYMBIQ 3.13 DUAL CHA 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK