SYMBIQ SCHANNEL 3.01

MDR report key: 3850887 · Received February 12, 2014

Report

Report Number
9615050-2014-01153
Event Type
Malfunction
Date Received
February 12, 2014
Date of Event
January 15, 2014
Report Date
January 17, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE ALARMED WITH A S233 (OVERTEMPERATURE) MALFUNCTION CODE. THIS WAS FOUND TO BE DUE TO THE FAN NOT ROTATING FAST ENOUGH TO KEEP THE PSC (POWER SUPPLY CONTROLLER) TEMPERATURE FROM EXCEEDING 70 DEGREES C. THE CAUSE OF THE OVERHEATING WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED DURING PREVENTATIVE MAINTENANCE AT THE USER FACILITY, THE DEVICE ALARMED WITH A "HIGH TEMP ALARM, MALFUNCTION TEMP TOO HIGH." THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93841 SYMBIQ SCHANNEL 3.01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA