FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ SCHANNEL 3.01
MDR report key: 3850887
·
Received February 12, 2014
Report
- Report Number
- 9615050-2014-01153
- Event Type
- Malfunction
- Date Received
- February 12, 2014
- Date of Event
- January 15, 2014
- Report Date
- January 17, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THE DEVICE ALARMED WITH A S233 (OVERTEMPERATURE) MALFUNCTION CODE. THIS WAS FOUND TO BE DUE TO THE FAN NOT ROTATING FAST ENOUGH TO KEEP THE PSC (POWER SUPPLY CONTROLLER) TEMPERATURE FROM EXCEEDING 70 DEGREES C. THE CAUSE OF THE OVERHEATING WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED DURING PREVENTATIVE MAINTENANCE AT THE USER FACILITY, THE DEVICE ALARMED WITH A "HIGH TEMP ALARM, MALFUNCTION TEMP TOO HIGH." THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93841 | SYMBIQ SCHANNEL 3.01 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |