FDA Adverse Event
Malfunction
Summary report: N
AFFINITY THREE BIRTHING BED
MDR report key: 3850831
·
Received March 21, 2014
Report
- Report Number
- 1824206-2014-00933
- Event Type
- Malfunction
- Date Received
- March 21, 2014
- Date of Event
- February 20, 2014
- Report Date
- February 20, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- PMA / PMN Number
- K915779
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE BRAKE/STEER CASTER NEED TO BE ADJUSTED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOW HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2007-2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN ADJUSTED THE BRAKE/STEER CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED IN L&D AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 169918 | AFFINITY THREE BIRTHING BED | BIRTHING BED | HDD | HILL-ROM, INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |