FDA Adverse Event
Malfunction
Summary report: N
AFFINITY BED
MDR report key: 3850823
·
Received March 21, 2014
Report
- Report Number
- 1824206-2014-00935
- Event Type
- Malfunction
- Date Received
- March 21, 2014
- Date of Event
- February 21, 2014
- Report Date
- February 21, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE FOOT END BRAKE/STEER CASTER WAS NOT HOLDING WHEN IN BRAKE MODE DUE TO A BROKEN LEFT CAM PIVOT. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE LEFT CAM PIVOT TO RESOLVE THE ISSUE BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE FRONT BRAKES WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN THE BED SHOP. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).