FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 3850823 · Received March 21, 2014

Report

Report Number
1824206-2014-00935
Event Type
Malfunction
Date Received
March 21, 2014
Date of Event
February 21, 2014
Report Date
February 21, 2014
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE FOOT END BRAKE/STEER CASTER WAS NOT HOLDING WHEN IN BRAKE MODE DUE TO A BROKEN LEFT CAM PIVOT. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE LEFT CAM PIVOT TO RESOLVE THE ISSUE BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE FRONT BRAKES WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN THE BED SHOP. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).