CLINITRON RITEHITE BED
Report
- Report Number
- 1824206-2014-00939
- Event Type
- Malfunction
- Date Received
- March 21, 2014
- Date of Event
- February 24, 2014
- Report Date
- February 24, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- INX
- PMA / PMN Number
- K964223
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
THE HILL-ROM TECHNICIAN FOUND THE CASTER NOT HOLDING DUE TO NORMAL WEAR AND TEAR. PER THE HILL-ROM USER MANUAL, WARNING: PATIENTS MAY USE THE BED FOR SUPPORT WHILE ENTERING AND EXITING; IF THE UNIT MOVES UNEXPECTEDLY, PATIENT INJURY COULD OCCUR. WHEN THE UNIT IS UNATTENDED, ENSURE THAT BOTH BRAKES ARE LOCKED. THE BRAKES FOR THE CLINITRON BED ARE LOCATED AT THE RIGHT, HEAD END AND THE LEFT, FOOT END OF THE UNIT. TO APPLY THE BRAKES, STEP ON THE LOWER END OF THE BRAKE LEVER TO LOCK THE WHEELS. TO RELEASE THE BRAKES, APPLY INWARD PRESSURE TO THE UPPER END OF THE BRAKE LEVER. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE BRAKE CASTERS WERE NO ENGAGING (BRAKE IS NOT HOLDING). THE BED WAS LOCATED IN A HILL-ROM SERVICE CENTER AND NOT IN USE. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 169506 | CLINITRON RITEHITE BED | BED, AIR FLUIDIZED | INX | HILL-ROM, INC. | 0800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |