FDA Adverse Event
Malfunction
Summary report: N
CORE UNIVERSAL DRIVER
MDR report key: 3850811
·
Received June 5, 2014
Report
- Report Number
- 0001811755-2014-02063
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 16, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON VISUAL INSPECTION, THE DEVICE WAS FOUND TO HAVE INTERNAL CORROSION. THE DEVICE WAS REPAIRED AND RETURNED TO THE USER FACILITY.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE CORE UNIVERSAL DRIVER CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 330037 | CORE UNIVERSAL DRIVER | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |