FDA Adverse Event Malfunction Summary report: N

CD HORIZON SPINAL SYSTEM

MDR report key: 3850766 · Received June 5, 2014

Report

Report Number
1030489-2014-02711
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 6, 2014
Report Date
May 7, 2014
Manufacturer
MDT SOFAMOR DANEK PUERTO RICO MFG
Product Code
KWP
PMA / PMN Number
K042025
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFO: VISUAL REVIEW CONFIRMS SCREW BREAKAGE APPROXIMATELY ~2-3 THREADS FROM BONE SCREW HEAD. OPTICAL EXAMINATION OF ADJACENT SURFACES AROUND THE AREA OF CRACK PROPAGATION DID NOT IDENTIFY MATERIAL SURFACE DEFECT THAT COULD CONTRIBUTE TO CRACK PROPAGATION. SIGNIFICANT DAMAGE AND/OR SMEARING NOTED TO ALL FRACTURE SURFACES. THE LOWER BROKEN PORTION OF THE BONE SCREW IS CONSISTENT WITH EXPLANTATION. FRACTURE SURFACE EXAMINATION IDENTIFIED A MULTI-MODAL FRACTURE, WITH PROGRESSIVE STRIATIONS EMANATING AWAY FROM THE ORIGIN OF INITIAL FRACTURE PROPAGATION AS NOTED, AS WELL AS INCREASED MATERIAL DISRUPTION, RIVER LINES, AND SHEAR LIPS, STARTING APPROXIMATELY ~50% OF THE WAY THROUGH THE CROSS-SECTIONAL AREA OF THE BONE SCREW, CONSISTENT WITH OVERLOAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A THREE LEVEL POSTERIOR LUMBAR FUSION. IT WAS REPORTED THAT THE SCREW BROKE IN HALF ABOUT 5MM BELOW THE TULIP HEAD. THE PATIENT UNDERWENT A REVISION SURGERY TO EXTEND THE CONSTRUCT DUE TO NON-UNION. NO ADDITIONAL INFORMATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330180 CD HORIZON SPINAL SYSTEM APPLIANCE, FIXATION, SPINAL INTERLAMINAL KWP MDT SOFAMOR DANEK PUERTO RICO MFG NA H12V0943

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention ROD, SET SCREW,