FDA Adverse Event Malfunction Summary report: N

CATGUT CHROMIC 2 (6) 50M

MDR report key: 3850465 · Received March 18, 2014

Report

Report Number
2916714-2014-00176
Event Type
Malfunction
Date Received
March 18, 2014
Report Date
March 18, 2014
Manufacturer
B. BRAUN SURGICAL S.A.
Product Code
GAL
PMA / PMN Number
K991223
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

US REPORTING AGENT NOTIFIED ON (B)(4) 2014. MFG SITE EVAL: SAMPLES RECEIVED: 1 CASSETTE. THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE/BATCH FOR THIS ISSUE. TESTED THE KNOT PULL TENSILE STRENGTH OF THE SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE OEM. ABSORPTION CONTROLS CONDUCTED ON THE SAMPLE WITH COLLAGENASE AFTER SEVERAL PERIODS OF TIME HAVE A RETENTION STRENGTH VALUES AS THE CURRENT ONES FOR THIS MATERIAL AND SIZE. THEREFORE, THIS THREAD DOES NOT DEGRADE QUICKER THAN OTHER BATCHES TESTED. REVIEWED THE BATCH MFG RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL OEM REQUIREMENTS. FINAL CONCLUSION: THE COMPLAINT IS NOT CORRESPONDING (NOT JUSTIFIED). ACTIONS ON PRODUCT: NOT APPLICABLE. CORRECTIVE/ PREVENTIVE ACTIONS: NOT APPLICABLE.

Description of Event or Problem · 1

COUNTRY OF COMPLAINT: (B)(6). THIS VETERINARY CUSTOMER COMPLAINS THE FOLLOWING: "THREE INCIDENCES WITH YIELDING/TRANSFER OF SUTURES ON AVERAGE FIVE DAYS AFTER USE". DEGRADATION OF SUTURE IS TOO QUICK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
161346 CATGUT CHROMIC 2 (6) 50M SUTURE GAL B. BRAUN SURGICAL S.A. F1154087 613112

Patients

Seq Age Sex Outcome Treatment
1 Other