FDA Adverse Event
Malfunction
Summary report: N
SILKAM BLACK 4/0 (1.5) 45CM HS18
MDR report key: 3850449
·
Received March 18, 2014
Report
- Report Number
- 2916714-2014-00140
- Event Type
- Malfunction
- Date Received
- March 18, 2014
- Report Date
- March 17, 2014
- Manufacturer
- B BRAUN SURGICAL S.A.
- Product Code
- GAP
- PMA / PMN Number
- K990089
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MFG SITE EVAL: SAMPLES RECEIVED - 1 CLOSED SAMPLE AND 1 OPEN SAMPLE WITH THE THREAD BROKEN (USED). THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE/BATCH. THERE ARE NO UNITS IN OEM STOCK. WE HAVE TESTED THE KNOT PULL TENSILE STRENGTH OF CLOSED SAMPLE RECEIVED AND THE RESULT FULFIL THE REQUIREMENTS OF THE OEM. REVIEWED THE BATCH MFG RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL OEM REQUIREMENTS. FINAL CONCLUSION: THE COMPLAINT IS NOT CORRESPONDING (NOT JUSTIFIED).
Description of Event or Problem · 1
COUNTRY OF COMPLAINT: (B)(6). THREAD CRACKED WHILE SUTURING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 159733 | SILKAM BLACK 4/0 (1.5) 45CM HS18 | SILK NON-ABSORBABLE SUTURE | GAP | B BRAUN SURGICAL S.A. | G0762563 | 612473V004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |