FDA Adverse Event Injury Summary report: N

OCTRODE

MDR report key: 3850291 · Received May 15, 2014

Report

Report Number
1627487-2014-26418
Event Type
Injury
Date Received
May 15, 2014
Date of Event
April 25, 2014
Report Date
April 25, 2014
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
P010032
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

IT WAS REPORTED THE PT WAS NOT RECEIVING STIMULATION TO HER BACK, HIPS AND LEGS. THE PT UNDERWENT SURGICAL INTERVENTION TO REPLACE THE LEADS. DURING THE PROCEDURE, IT WAS REVEALED THE LEADS WERE COILED TOGETHER IN A POCKET AND NOT DOWN TO THE IPG. IT WAS NOTED THAT THE ANCHORS WERE LOOSE AND NOT SECURED TO THE LEADS. BOTH LEADS WERE EXPLANTED BUT THE PHYSICIAN WAS ONLY ABLE TO IMPLANT ONE NEW LEAD. STRAIN RELIEF LOOPS WERE USED AND THE ANCHORS WERE SUTURED SECURELY. POSTOPERATIVELY THE PT WAS REPROGRAMMED AND SHE WAS RECEIVING EFFECTIVE STIMULATION. THE LEADS WILL NOT BE RETURNED TO THE MANUFACTURER FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291464 OCTRODE SCS LEAD GZB ST. JUDE MEDICAL - NEUROMODULATION 3186 4349932

Patients

Seq Age Sex Outcome Treatment
1 74 YR IMPLANTED:| SCS IPG: MODEL 3716