FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3850076 · Received June 4, 2014

Report

Report Number
2531779-2014-15786
Event Type
Malfunction
Date Received
June 4, 2014
Report Date
May 29, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/15/2014 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT AN UNEXPLAINED POWER REBOOT WAS RECORDED ON (B)(4) 2014. VISUAL INSPECTION REVEALED THAT THE BATTERY COMPARTMENT WAS INTACT; NO DAMAGE WAS OBSERVED. THE RETURNED BATTERY CAP WAS UNDAMAGED AND WAS ABLE TO BE FULLY SECURED TO THE PUMP. A 24 HOUR DURATION TEST WAS PERFORMED USING THE RETURNED BATTERY CAP; NO POWER INTERRUPTIONS OCCURRED. THE BATTERY CAP CONTACT HEIGHT AND WIDTH MEASUREMENTS WERE FOUND TO BE WITHIN THE REQUIRED SPECIFICATIONS. THE PUMP CASE WAS REMOVED AND NO EVIDENCE OF INTERNAL MOISTURE OR DAMAGE TO THE POWER CIRCUIT WAS OBSERVED. THE COMPLAINT OF THE PUMP LOSING POWER INTERMITTENTLY WAS OBSERVED IN THE PUMP HISTORY BUT WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. IT WAS REPORTED THAT THE PUMP WAS LOSING POWER INTERMITTENTLY. IT WAS NOTED THAT THE POWER ISSUE OCCURRED DURING OR IMMEDIATELY AFTER A BATTERY CHANGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
326208 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 35 YR