FDA Adverse Event Malfunction Summary report: N

MEDTRONIC

MDR report key: 3849564 · Received June 4, 2014

Report

Report Number
3849564
Event Type
Malfunction
Date Received
June 4, 2014
Date of Event
May 16, 2014
Report Date
May 22, 2014
Product Code
JXG
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TX, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING INSERTION ON (B)(6) 2014, THERE WAS SHEARING OFF OF THE... THE FRAGMENT WAS RETRIEVED ON (B)(6) 2014. WHILE USING FLUOROSCOPY FOR GUIDANCE, THE TIP OF A LARGE CALIBER SPINAL NEEDLE WAS PLACED WITHIN THE THECAL SAC VIA A LEFT L2/L3 INTERLAMINAR APPROACH. CSF WAS OBTAINED. INITIALLY LUMBAR DRAIN WAS INSERTED THROUGH THE SPINAL NEEDLE AND EXTENDED SUPERIORLY TO APPROX THE T10 LEVEL, BUT WOULD NOT ADVANCE FURTHER; THE CATHETER SEEMED TO TURN BACK ON ITSELF; IT WAS AT THAT INSTANT THAT THE WIRE TO THE DRAIN WAS VISIBLE OUTSIDE THE NEEDLE. THIS CATHETER WAS REMOVED. THE SECOND LUMBAR DRAIN WAS INSERTED THROUGH THE SPINAL NEEDLE AND WAS ABLE TO BE SUCCESSFULLY TRAVERSED SUPERIORLY UP THE THECAL SAC TO THE MID T7 LEVEL. SPINAL NEEDLE WAS REMOVED. PT TOLERATED PROCEDURE WELL. AT THE END OF THE PROCEDURE APPROX 15CM OF THE INITIAL LUMBAR DRAIN WAS NOTED TO BE SHEARED OFF AND LIKELY LOCATED WITHIN THE PT'S THECAL SAC. CT SCAN SHOWED LUMBAR DRAIN ENTERING AT L2-L3 LEVEL AND TIP AT L7. A FRAGMENTED LUMBAR DRAIN WAS SEEN AT THE T10 TO L3 LEVEL ON THE RIGHT. ON (B)(6) 2014, PT HAD SURGERY TO REMOVE THE INTRATHECAL CATHETER FRAGMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
327112 MEDTRONIC LUMBAR CATHETER JXG D58054 (PROBABLE)

Patients

Seq Age Sex Outcome Treatment
1 28 YR