FDA Adverse Event Malfunction Summary report: N

REMB SAG SAW

MDR report key: 3845045 · Received June 3, 2014

Report

Report Number
0001811755-2014-02025
Event Type
Malfunction
Date Received
June 3, 2014
Date of Event
May 9, 2014
Report Date
May 9, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT, BIAS CURRENT ERROR ON THE CONSOLE, WAS DUPLICATED. UPON DISASSEMBLY FOR VISUAL INSPECTION THE SERVICE TECHNICIAN OBSERVED INTERNAL MOTOR ASSEMBLY CORROSION.

Description of Event or Problem · 1

THE REMB SAGITTAL SAW WAS RETURNED TO STRYKER INSTRUMENTS FOR SERVICE. DURING FUNCTIONAL TESTING BY A SERVICE TECHNICIAN, IT WAS FOUND THAT THE DEVICE CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.