LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-00796
- Event Type
- Malfunction
- Date Received
- March 12, 2014
- Date of Event
- February 12, 2014
- Report Date
- March 11, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE ELECTRODE BELT FAILED INCOMING FUNCTIONALITY TESTING. THE CAUSE FOR THE FAILURE WAS ISOLATED TO OPEN CONNECTIONS ON THE DISTRIBUTION NODE PCA BOARD. THE GREEN, BLACK, RED, AND WHITE WIRES WERE BROKEN FROM THEIR SOLDER JOINT ALONG THE DN TO REAR THERAPY ELECTRODE CABLE. THE ROOT CAUSE FOR THE OPEN CONNECTIONS COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, ELECTRODE BELT SN (B)(4) FAILED INCOMING FUNCTIONALITY TESTING. THE LAST PATIENT USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149513 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |