LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-00776
- Event Type
- Malfunction
- Date Received
- March 12, 2014
- Date of Event
- February 11, 2014
- Report Date
- March 10, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. UPON INVESTIGATION THE CHARGER WAS RESETTING. THE CAUSE OF THE RESETS WAS ISOLATED TO THE CHARGER/MODEM'S CA BOARD. THE ROOT CAUSE FOR THE RESETS WAS UNABLE TO BE POSITIVELY DETERMINED, BUT WAS LIKELY RELATED TO AN INTERMITTENT BGA CONNECTION ON THE CA BOARD. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE CHARGER / MODEM. THE LAST PATIENT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.
REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, CHARGER/MODEM SN (B)(4) WAS RESETTING. THE LAST PATIENT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149296 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |