FDA Adverse Event Malfunction Summary report: N

CARE ASSIST BED

MDR report key: 3844778 · Received March 12, 2014

Report

Report Number
9615739-2014-00105
Event Type
Malfunction
Date Received
March 12, 2014
Date of Event
February 12, 2014
Report Date
February 12, 2014
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE BRAKE CASTERS HAD DAMAGED STEMS WHICH ALSO DAMAGED THE BRAKE CASTER SUPPORTS. AN INVESTIGATION WAS PERFORMED AND THIS ISSUE WAS ADDRESSED UNDER AN ENGINEERING CHANGE. THE TECHNICIAN REPLACED THE BRAKE CASTER, BRAKE STEER CASTER AND SUPPORTS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECHNICIAN REPORTED THE BRAKE CASTERS DO NOT HOLD. THE BED WAS LOCATED IN THE BED REPAIR AREA AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149302 CARE ASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1170