FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3844591 · Received June 3, 2014

Report

Report Number
2939301-2014-13165
Event Type
Malfunction
Date Received
June 3, 2014
Report Date
May 26, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (07/16/2014).THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 6/10/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 7/4/2014 WITH THE FOLLOWING FINDINGS:THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE RETURNED TEST STRIPS WERE FOUND TO HAVE SCRATCHED ELECTRODES WITH NO ASSIGNABLE CAUSE. IN ADDITION, A SECONDARY ISSUE WAS NOTED WHEN THE ELECTRODES ON THE TEST STRIP ARE SCORED CAUSING A SHORT-ERROR 2. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING DOES NOT TURN ON - STRIP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323957 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL

Patients

Seq Age Sex Outcome Treatment
1 66 YR