OT VERIO IQ METER
Report
- Report Number
- 2939301-2014-13165
- Event Type
- Malfunction
- Date Received
- June 3, 2014
- Report Date
- May 26, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (07/16/2014).THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 6/10/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 7/4/2014 WITH THE FOLLOWING FINDINGS:THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE RETURNED TEST STRIPS WERE FOUND TO HAVE SCRATCHED ELECTRODES WITH NO ASSIGNABLE CAUSE. IN ADDITION, A SECONDARY ISSUE WAS NOTED WHEN THE ELECTRODES ON THE TEST STRIP ARE SCORED CAUSING A SHORT-ERROR 2. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING DOES NOT TURN ON - STRIP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 323957 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |