FDA Adverse Event Death Summary report: N

AMNISURE

MDR report key: 3844576 · Received May 28, 2014

Report

Report Number
3005345832-2014-00003
Event Type
Death
Date Received
May 28, 2014
Report Date
May 22, 2014
Manufacturer
AMNISURE INTERNATIONAL, LLC
Product Code
NQM
PMA / PMN Number
K081767
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY HAS EVALUATED THIS EVENT AND HAS DETERMINED THAT THE RESULT OBTAINED FROM THE DEVICE DID NOT CONTRIBUTE OR CAUSE TO THE DEATH OF THE FETUS. HOWEVER, THE COMPANY HAS DECIDED TO SUBMIT THIS MDR FOR INFO PURPOSES IN AN ABUNDANCE OF CAUTION AND TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803. THE COMPANY WILL SUPPLEMENT THIS MDR WITH ADDITIONAL INFO, AS NECESSARY. BASED ON THE INFO PROVIDED, IT IS CLEAR THAT THE TEST WAS NOT PERFORMED ACCORDING TO THE PRODUCT INSTRUCTIONS FOR USE. THE HEALTHCARE FACILITY REPORTED UNEXPECTED NEGATIVE RESULTS ALTHOUGH THE PT FIRST REPORTED SIGNS/SYMPTOMS OF RUPTURE 24 HOURS BEFORE BEING TESTED. THE AMNISURE PRODUCT INSTRUCTIONS FOR USE STATES THAT A FALSE NEGATIVE MAY OCCUR WHEN A SAMPLE IS TAKEN 12 HOURS OF LATER AFTER A RUPTURE DUE TO OBSTRUCTION OF THE RUPTURE BY THE FETUS A RUPTURE DUE TO OBSTRUCTION OF THE RUPTURE BY THE FETUS OR RESEALING OF THE AMNIOTIC SAC. IT SHOULD BE NOTED THAT THE PT ALSO PRESENTED WITH OTHER COMPLICATIONS. IN GENERAL, THE PRODUCT INSTRUCTION FOR USE ADVISES PHYSICIANS TO EVALUATE ALL RESULTS IN CONJUNCTION WITH ADDITIONAL CLINICAL INFO.

Description of Event or Problem · 1

A COMPLAINT WAS RECEIVED ON (B)(4) 2013 REGARDING A PT WHO WAS AT (B)(6) GESTATION AND HAD FIRST REPORTED SIGNS/SYMPTOMS OF RUPTURE 24 HOURS BEFORE BEING TESTED, AND WAS NOT REPORTING CONTINUOUS LEAKAGE. THE PT'S AMNIOTIC FLUID INDEX WAS 22 AND SHE TESTED NEGATIVE WITH NITRAZINE AND AMNISURE. CLINICAL STAFF NOTICED A CREAMY DISCHARGE FROM THE PT'S VAGINAL ORIFICE. PT WAS SENT HOME AT AROUND 10:45AM. PT RETURNED TO HOSPITAL 8 HOURS (5:30 PM) LATER AND TESTED POSITIVE WITH AMNISURE AND THE CLINICAL DECISION WAS TO DISCHARGE AND SEND HOME. PT RETURNED THE NEXT DAY AND DISPLAYED SYMPTOMS OF OBVIOUS ROM AND HAD DEVELOPED A FEVER. PT WAS INDUCED AND SHE DELIVERED A STILLBORN FETUS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
314796 AMNISURE RAPID FETAL MEMBRANE RUPTURE TEST NQM AMNISURE INTERNATIONAL, LLC FMRT-1 A2021A

Patients

Seq Age Sex Outcome Treatment
1 Death