FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3844484 · Received June 2, 2014

Report

Report Number
2531779-2014-15583
Event Type
Malfunction
Date Received
June 2, 2014
Report Date
May 28, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

PRODUCT ANALYSIS: THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/10/2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT COULD NOT BE DUPLICATED OR CONFIRMED WITH INVESTIGATION. REVIEW OF THE BLACK BOX REVEALED DATA FROM THE DATE OF COMPLAINT HAD BEEN OVERWRITTEN DUE TO CONTINUED PUMP USE; NO EVIDENCE OF A SHORTENED BATTERY LIFE OF EXCESSIVE POWER DRAW WAS OBSERVED. NO DAMAGE WAS OBSERVED TO THE BATTERY COMPARTMENT; THE BATTERY CAP WAS ABLE TO BE PROPERLY SECURED TO THE PUMP. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE WITHOUT ISSUE OR ALARM. THE PUMP¿S POWER DRAWS WERE OBSERVED TO BE WITHIN SPECIFICATION. THERE WAS NO EVIDENCE OF DAMAGE OR DEFECT TO THE PUMP¿S INTERNAL COMPONENTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321563 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 20 YR