FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPSINSTRUMENT

MDR report key: 3844270 · Received June 2, 2014

Report

Report Number
2955842-2014-03394
Event Type
Malfunction
Date Received
June 2, 2014
Date of Event
May 1, 2014
Report Date
May 5, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THE PITCH UP CABLE BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT WRIST ARE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE BROKEN CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE CABLE BROKE ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321926 FENESTRATED BIPOLAR FORCEPSINSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10130530 117

Patients

Seq Age Sex Outcome Treatment
1 66 YR