FDA Adverse Event Injury Summary report: N

X3 TRIATHLON CS INS SIZE4 13MM

MDR report key: 3844144 · Received June 2, 2014

Report

Report Number
0002249697-2014-01989
Event Type
Injury
Date Received
June 2, 2014
Date of Event
May 6, 2014
Report Date
May 6, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JWH
PMA / PMN Number
K063423
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IT WAS NOTED THAT THE DEVICE IS NOT AVAILABLE FOR EVALUATION DUE TO HOSPITAL POLICY. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF RECEIVED, WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT. NOT RETURNED TO MANUFACTURER.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING SEATING LOCKING ISSUES INVOLVING A X3 TRIATHLON CS INS SIZE4 13MM WAS REPORTED. THE EVENT WAS CONFIRMED. DEVICE EVALUATION NOT PERFORMED AS NO DEVICES WERE RETURNED FOR EVALUATION. MEDICAL RECORDS AND EVALUATION INDICATED THAT INSUFFICIENT MEDICAL INFORMATION WAS PROVIDED. THE CLINICIAN DID STATE THAT THE X-RAYS SHOW THE LOCKING MECHANISM DISLODGED. DEVICE HISTORY REVIEW INDICATED THAT THE SPECIFIED LOT WAS ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW INDICATED THAT THERE HAVE NOT BEEN ANY OTHER EVENTS FOR THE SPECIFIED LOT. THE REPORTED EVENT INVOLVES AN ISSUE WITH THE LOCKING MECHANISM FOR THE POLY INSERT. THE PATIENT RECEIVED A POLY SWAP THE NEXT DAY. THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE NOT ENOUGH INFORMATION WAS PROVIDED. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME.

Description of Event or Problem · 1

EVERYTHING WENT ACCORDING TO PLAN IN SURGERY AND NOTHING SEEMED ABNORMAL. THAT AFTERNOON DOCTOR INDICATED TO THE REP THAT THERE MAY BE LOCKING MECHANISM ISSUE WITH THE POLY. PATIENT WAS SCHEDULED FOR POLY SWAP ON THE (B)(6) 2014.

Description of Event or Problem · 1

EVERYTHING WENT ACCORDING TO PLAN IN SURGERY AND NOTHING SEEMED ABNORMAL. THAT AFTERNOON DOCTOR INDICATED TO THE REP THAT THERE MAY BE LOCKING MECHANISM ISSUE WITH THE POLY. PATIENT WAS SCHEDULED FOR POLY SWAP ON THE (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322406 X3 TRIATHLON CS INS SIZE4 13MM IMPLANT JWH STRYKER ORTHOPAEDICS-MAHWAH LDS546

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention