ROTATING MULTIPLE CLIP APPLIER
Report
- Report Number
- 3005075853-2014-03723
- Event Type
- Malfunction
- Date Received
- June 2, 2014
- Report Date
- April 7, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). RESULTS: JAW MISALIGNED, SCISSORED CLIP. THE ANALYSIS RESULTS OF THE ER420 DEVICE FOUND THAT IT WAS RECEIVED WITH THE JAWS MISALIGNED AND A CLIP IN JAWS. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE CLIPS WERE FED AS INTENDED; HOWEVER, DUE TO THE MISALIGNED CONDITION OF THE JAWS 16 SCISSORED CLIPS WERE FORMED. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE TWICE ONLY THREE CLIPS AND ONCE DID NOT FUNCTION AT ALL. NO FURTHER INFORMATION IS AVAILABLE. NO PATIENT CONSEQUENCES REPORTED. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322002 | ROTATING MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | K4CC1V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |