FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 3844129 · Received June 2, 2014

Report

Report Number
3005075853-2014-03723
Event Type
Malfunction
Date Received
June 2, 2014
Report Date
April 7, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: JAW MISALIGNED, SCISSORED CLIP. THE ANALYSIS RESULTS OF THE ER420 DEVICE FOUND THAT IT WAS RECEIVED WITH THE JAWS MISALIGNED AND A CLIP IN JAWS. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE CLIPS WERE FED AS INTENDED; HOWEVER, DUE TO THE MISALIGNED CONDITION OF THE JAWS 16 SCISSORED CLIPS WERE FORMED. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE TWICE ONLY THREE CLIPS AND ONCE DID NOT FUNCTION AT ALL. NO FURTHER INFORMATION IS AVAILABLE. NO PATIENT CONSEQUENCES REPORTED. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322002 ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA K4CC1V

Patients

Seq Age Sex Outcome Treatment
1