MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2014-03376
- Event Type
- Malfunction
- Date Received
- June 2, 2014
- Date of Event
- March 20, 2014
- Report Date
- May 19, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION OBSERVED THAT THE INSTRUMENT'S BLADES FAILED THE CUT TEST, BUT THE BLADES WERE NOT DAMAGED. THE INSTRUMENT WAS PLACED ON AN IN-HOUSE DA VINCI SYSTEM FOR LATEX CUT TEST AND THE SCISSORS DID NOT CUT CLEANLY THROUGH (B)(4) LATEX. THE LATEX GOT SNAGGED AT THE SCISSOR TIPS. THE BLADE EDGES WERE NOT DAMAGED, BUT EDGES EXHIBIT WEAR AT THE TIPS, WHICH NEGATIVELY AFFECTED CUT PERFORMANCE. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT'S TUBE EXTENSION WAS CRACKED. THE TUBE EXTENSION WAS CRACKED AT ONE OF THE AXIAL KEYS, THERE WERE NO MISSING MATERIALS AND NO BROKEN PIECES WAS FOUND. EVIDENCE NOT CONCLUSIVE, BUT CRACKED DAMAGE MAY BE DUE TO LIKELY MISHANDLING/MISUSE. AN ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT'S MAIN TUBE EXHIBITED SCRATCH MARKS/ABRASIONS. THE DISTAL END OF MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. AN ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT FLUSH TUBE WAS KINKED. THE HOUSING WAS REMOVED TO FIND FLUSH TUBE WITH A COUPLE OF KINKS IN THE BACKEND. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT, IN A FRAGMENT-CAUSING FAILURE (SUCH AS A CABLE FAILURE, A CRIMP SEPARATING FROM A CABLE ETC.) FOR THE MCS INSTRUMENT THE TIP COVER ACCESSORY WILL HOLD SUCH FRAGMENTS INSIDE THE TIP COVER, PREVENTING LOSS INTO THE PATIENT DURING THE REMOVAL OF THE INSTRUMENT. BECAUSE THE TIP COVER ACTS AS A BARRIER, THESE FAILURES, WHICH MIGHT BE REPORTABLE IN OTHER INSTRUMENTS, ARE NOT REPORTABLE EVENTS FOR THE MCS INSTRUMENT; HOWEVER, THE MAIN TUBE DAMAGE, WHICH WAS FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, THE CUSTOMER NOTED THAT THE MONOPOLAR CURVED SCISSORS INSTRUMENT WOULD NOT WORK ANYMORE. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321448 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-12 | M10130617 108 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |