LARGE NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2014-03367
- Event Type
- Malfunction
- Date Received
- June 2, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 16, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION OBSERVED THAT ONE GRIP CABLE WAS FOUND TO BE BROKEN AT THE DISTAL IDLER. THE IDLER PULLEY SPUN FREELY AND DID NOT EXHIBIT ANY DAMAGE. THE CABLE SEGMENT STUCK OUT AT THE WRIST. OTHER CABLES AT WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S GRIP CABLE COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT DURING REPROCESSING OF THE LARGE NEEDLE DRIVER INSTRUMENT, IT WAS NOTED THAT THE CABLE ON THE INSTRUMENT WAS FRAYED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321418 | LARGE NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420006-08 | M11131213 111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |