FDA Adverse Event Malfunction Summary report: N

SPEEDBAND SUPERVIEW SUPER 7?

MDR report key: 3843816 · Received June 2, 2014

Report

Report Number
3005099803-2014-02066
Event Type
Malfunction
Date Received
June 2, 2014
Date of Event
May 10, 2014
Report Date
May 11, 2014
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
FHN
PMA / PMN Number
K020824
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

REPORTED EVENT OF BANDS FAILED TO DEPLOY. ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.

Additional Manufacturer Narrative · 1

THE LIGATOR HEAD WAS THE ONLY COMPONENT RETURNED. A VISUAL EXAMINATION OF THE RETURNED COMPONENT NOTED THAT THE SUTURE WAS INTACT AND REMAINED ATTACHED TO THE HEAD. AN EXAMINATION FOUND THAT ALL SEVEN BANDS WERE PRESENT ON THE LIGATOR HEAD WITH SIX OF THE BANDS MOVED OUT OF POSITION. THE LIGATOR HEAD TEETH PRESENTED DAMAGE GIVEN THE EVENT DESCRIPTION AND CONDITION OF THE RETURNED DEVICE, THERE ISN'T ENOUGH INFORMATION TO DETERMINE A PROBABLE ROOT CAUSE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPEEDBAND SUPERVIEW SUPER 7 DEVICE WAS USED IN THE ESOPHAGUS DURING AN ENDOSCOPIC VARICEAL LIGATION PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BAND FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPEEDBAND SUPERVIEW SUPER 7 DEVICE WAS USED IN THE ESOPHAGUS DURING AN ENDOSCOPIC VARICEAL LIGATION PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BAND FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321650 SPEEDBAND SUPERVIEW SUPER 7? LIGATOR, HEMORRHOIDAL FHN BOSTON SCIENTIFIC - SPENCER M00542251 16529249

Patients

Seq Age Sex Outcome Treatment
1 54 YR