CENTURY BED
Report
- Report Number
- 1824206-2014-00384
- Event Type
- Malfunction
- Date Received
- February 12, 2014
- Date of Event
- January 14, 2014
- Report Date
- January 14, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
UPON INVESTIGATION, THE TECHNICIAN FOUND THAT A PREVIOUS SERVICING COMPANY HAD REPLACED A ROLL PIN WITH A WIRE AND TWO SCREWS ON A PREVIOUS REPAIR. PER THE HILL-ROM SERVICE MANUAL THE CENTURY BED REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM ANNUAL PREVENTATIVE MAINTENANCE. ENSURE THE BED DOES NOT MOVE WHEN THE BRAKES ARE ACTIVATED. ADJUST THE BRAKES IF NECESSARY. INSPECT AND ADJUST THE STEERING MECHANISM IF NECESSARY. CHECK THE CASTER TIRES FOR CUTS, WEAR, TREAT, ETC. REPLACE IF NECESSARY. THE DEVICE REFERENCED IN THIS REPORT IS OVER 12 MONTHS OLD AND ACCORDING TO THE RECOMMENDED PREVENTIVE MAINTENANCE SCHEDULE AT LEAST ONE PM SHOULD BE PERFORMED EVERY 12 MONTHS. THEREFORE, THE ACCEPTANCE CRITERIA, BASED ON THE DHR, WERE DETERMINED TO BE ACCEPTABLE AT THE TIME THE DEVICE WAS RELEASED. GIVEN THE TIME WHICH HAS ELAPSED, THE ALLEGED FAILURE OF THE DEVICE WOULD NOT BE RELATED TO ANY ISSUE WHICH OCCURRED DURING THE MFG PROCESS. THE TECHNICIAN REPLACED THE ROLL PIN TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
THE TECHNICIAN REPORTED THAT WHEN THE FOOT END BRAKE PEDAL IS SET TO BRAKE MODE THE BRAKE CASTERS DO NOT HOLD. THE BED WAS LOCATED IN THE PARKING GARAGE OF THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 92821 | CENTURY BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1400 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |