TRANSTAR STRETCHER
Report
- Report Number
- 3006697241-2014-00139
- Event Type
- Malfunction
- Date Received
- February 12, 2014
- Date of Event
- January 15, 2014
- Report Date
- January 15, 2014
- Manufacturer
- HILLROM DE MEXICO S DE RL DE CV
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
THE TECHNICIAN FOUND THE SIDE RAIL IS MISSING THE SHOULDER BOLT CAUSING IT NOT TO LATCH. PER THE HILL-ROM SERVICE MANUAL IT IS NECESSARY FOR STRETCHERS TO HAVE AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU DO ANNUAL PREVENTATIVE MAINTENANCE. MAKE SURE THE SIDE RAILS ARE NOT BENT OR TWISTED. MAKE SURE THE LATCH MECHANISMS OPERATE CORRECTLY. YOU MUST HEAR AN AUDIBLE CLICK WHEN THE SIDE RAIL IS RAISED TO THE UP POSITION. CHECK FOR LOOSE OR MISSING HARDWARE. TIGHTEN, REPAIR OR REPLACE AS NECESSARY. THE TECHNICIAN REPLACED THE SIDE RAIL SHOULDER BOLT TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
PER THE HILL-ROM SERVICE MANUAL IT IS NECESSARY FOR STRETCHERS TO HAVE AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU DO ANNUAL PREVENTATIVE MAINTENANCE. MAKE SURE THE SIDE RAILS ARE NOT BENT OR TWISTED. MAKE SURE THE LATCH MECHANISMS OPERATE CORRECTLY. YOU MUST HEAR AN AUDIBLE CLICK WHEN THE SIDE RAIL IS RAISED TO THE UP POSITION. CHECK FOR LOOSE OR MISSING HARDWARE. TIGHTEN, REPAIR OR REPLACE AS NECESSARY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 92924 | TRANSTAR STRETCHER | WHEELED STRETCHER | FPO | HILLROM DE MEXICO S DE RL DE CV | 8000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |