HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-17657
- Event Type
- Malfunction
- Date Received
- June 2, 2014
- Date of Event
- January 26, 2014
- Report Date
- May 8, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EZ
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN INCREASED INTRA PERITONEAL VOLUME (IIPV) EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE ROOT CAUSE COULD NOT BE IDENTIFIED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 112 (NIGHT DRAIN #12) ALARM WAS IDENTIFIED IN THE LOG. THIS ALARM INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE, INDICATING AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE ALARM OCCURRED ON (B)(6) 2014 AT 07:45:51. NO ADDITIONAL INFORMATION IS AVAILABLE.