FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BARIATRIC PLUS BED
MDR report key: 3843727
·
Received February 12, 2014
Report
- Report Number
- 1824206-2014-00398
- Event Type
- Malfunction
- Date Received
- February 12, 2014
- Date of Event
- January 15, 2014
- Report Date
- January 15, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- PMA / PMN Number
- K122473
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON INVESTIGATION, THE TECHNICIAN FOUND THAT THE MATTRESS ON THE BED WAS A LITTLE DISPLACED BY THE WEIGHT OF THE PATIENT. THE ACCOUNT REPOSITIONED THE PATIENT AND THE SIDE RAIL LATCHED PROPERLY, USER ERROR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN STATED THE SIDE RAIL FUNCTIONS AS DESIGNED AND NO REPAIR WAS NEEDED. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
THE TECHNICIAN STATED THE ACCOUNT REPORTED THE SIDE RAIL LATCH IS BROKEN. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT NUMBER (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 92822 | TOTALCARE BARIATRIC PLUS BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |