FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC PLUS BED

MDR report key: 3843727 · Received February 12, 2014

Report

Report Number
1824206-2014-00398
Event Type
Malfunction
Date Received
February 12, 2014
Date of Event
January 15, 2014
Report Date
January 15, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON INVESTIGATION, THE TECHNICIAN FOUND THAT THE MATTRESS ON THE BED WAS A LITTLE DISPLACED BY THE WEIGHT OF THE PATIENT. THE ACCOUNT REPOSITIONED THE PATIENT AND THE SIDE RAIL LATCHED PROPERLY, USER ERROR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN STATED THE SIDE RAIL FUNCTIONS AS DESIGNED AND NO REPAIR WAS NEEDED. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECHNICIAN STATED THE ACCOUNT REPORTED THE SIDE RAIL LATCH IS BROKEN. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT NUMBER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
92822 TOTALCARE BARIATRIC PLUS BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1