CENTRA BED
Report
- Report Number
- 1824206-2014-00386
- Event Type
- Malfunction
- Date Received
- February 12, 2014
- Date of Event
- January 14, 2014
- Report Date
- January 14, 2014
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
UPON INVESTIGATION, TECHNICIAN FOUND THE CASTERS ON THE DEVICE WERE WORN WITH WEAR AND TEAR ASSOCIATED WITH AGE. THE CASTERS ON THE DEVICE HAVE BEEN IN SERVICE SINCE 1991 AND THERE IS NO RECORD OF THE CASTERS BEING REPLACED BEFORE THIS INCIDENT. THE DEVICE REFERENCED IN THIS REPORT IS OVER 12 MONTHS OLD AND ACCORDING TO THE RECOMMENDED PREVENTIVE MAINTENANCE SCHEDULE AT LEAST ONE PM SHOULD BE PERFORMED EVERY 12 MONTHS. THEREFORE, THE ACCEPTANCE CRITERIA, BASED ON THE DHR, WERE DETERMINED TO BE ACCEPTABLE AT THE TIME THE DEVICE WAS RELEASED. GIVEN THE TIME WHICH HAS ELAPSED, THE ALLEGED FAILURE OF THE DEVICE WOULD NOT BE RELATED TO ANY ISSUE WHICH OCCURRED DURING THE MANUFACTURING PROCESS. THE TECHNICIAN REPLACED THE BRAKE CASTERS, BRAKE STEER CASTER AND THE BRAKE MECHANISM ASSEMBLY TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
THE TECHNICIAN REPORTED HE FOUND THE BED VACANT AND IT HAD A NOTE ON IT STATING THE BRAKES ARE NOT FUNCTIONING. THE BED WAS LOCATED IN THE BASEMENT OF THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. HIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT NUMBER (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93341 | CENTRA BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 852 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |