ZENITH FLEX AAA ENDOVASCULAR GRAFT ILIAC LEG
Report
- Report Number
- 1820334-2014-00222
- Event Type
- Injury
- Date Received
- May 28, 2014
- Date of Event
- April 30, 2014
- Report Date
- May 1, 2014
- Manufacturer
- COOK, INC.
- Product Code
- MIH
- PMA / PMN Number
- P020018
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVENT IS STILL UNDER INVESTIGATION.
ON (B)(6) 2012, A (B)(6) FEMALE PATIENT HAD UNDERGONE AAA REPAIR WITH ZENITH FLEX WITH LEFT APPROACH. THE PATIENT WAS SUITABLE FOR ENDOVASCULAR REPAIR AND THE PROCEDURE WAS CONDUCTED AS LABELED. TYPE 1A ENDOLEAK WAS CONFIRMED, SO AN EXTENSION WAS ADDITIONALLY PLACED. THE PROCEDURE WAS FINISHED AFTER CONFIRMING THE TYPE 1 ENDOLEAK WAS RESOLVED. ACCORDING TO ADDITIONAL INFORMATION PROVIDED ON (B)(6) 2014, THE CONTRALATERAL ILIAC LEG GRAFT WAS PLACED WITH SHORT OVERLAP INSIDE THE CONTRALATERAL LIMB OF THE MAIN BODY. ON (B)(6) 2012, FOLLOW-UP CT ANGIO DID NOT CONFIRM ANEURYSM ENLARGEMENT. ON (B)(6) 2013, FOLLOW-UP CT ANGIO DID NOT CONFIRM ANEURYSM ENLARGEMENT. SINCE THE PATIENT FELT SICK WITH CONTRAST MEDIUM, FOLLOW-UP SIMPLE CT WERE PERFORMED SEVERAL TIMES OTHER THAN CT ANGIO DESCRIBED ABOVE. ON UNKNOWN DATE, A SIMPLE FOLLOW-UP CT SHOWED ANEURYSM ENLARGEMENT ACCORDING TO A RADIOLOGIST'S COMMENT. WITH THE RADIOLOGIST'S COMMENT, CT ANGIO WAS PERFORMED AS FOLLOW-UP. HOWEVER, NO SEVERE ANEURYSM ENLARGEMENT WAS OBSERVED. THE PHYSICIAN DECIDED TO CONTINUE THE FOLLOW-UP. ON 04/30/2014 MORNING, URGENT PROCEDURE WAS CONDUCTED DUE TO AAA RUPTURE. ANGIOGRAPHY CONFIRMED ENDOLEAK FROM JUNCTION PART BETWEEN CONTRALATERAL ILIAC LEG GRAFT AND MAINBODY. THEN, ANOTHER MANUFACTURER'S STENT GRAFT WAS ADDITIONALLY PLACED OVER THE CIA AND EIA COVERING THE PLACE WHERE ENDOLEAK WAS CONFIRMED. CONFIRMATORY ANGIOGRAPHY WAS PERFORMED SEVERAL TIMES AFTER THE ADDITIONAL PLACEMENT, AND THE PROCEDURE WAS FINISHED AFTER CONFIRMING THAT THE ENDOLEAK WAS RESOLVED. THE PATIENT REMAINS HOSPITALIZED. LOWERING OF BLOOD PRESSURE IS OBSERVED AS OF (B)(6) 2014. AS OF (B)(6) 2014, BLOOD PRESSURE IS STABILIZED THOUGH, WE CANNOT BE OPTIMISTIC ABOUT PATIENT'S CONDITIONS; RENAL DYSFUNCTION IS OBSERVED. IT IS POSSIBLE THAT LIVER DYSFUNCTION AND/OR INFECTIOUS DISEASE WILL BE OBSERVED HEREAFTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314638 | ZENITH FLEX AAA ENDOVASCULAR GRAFT ILIAC LEG | MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | COOK, INC. | NA | F2707502 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization |