FDA Adverse Event
Malfunction
Summary report: N
PUMP IN STYLE (ITEM # UNK)
MDR report key: 3843571
·
Received March 25, 2014
Report
- Report Number
- 1419937-2014-00295
- Event Type
- Malfunction
- Date Received
- March 25, 2014
- Date of Event
- March 10, 2014
- Report Date
- March 10, 2014
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER'S REPORT OF A SPARK COMING FROM THE HOUSING OF HER POWER ADAPTER IS A SAFETY ISSUE. IN F/U WITH THE CUSTOMER, THE CUSTOMER STATED THAT THERE WERE NO INJURIES THAT RESULTED AND THAT SHE DISCARDED THE POWER ADAPTER. AS A RESULT OF CAPA (B)(4) WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES, A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. ACTIVITIES RELATED TO THIS NOTIFICATION ARE ON-GOING.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT WHEN SHE WENT TO UNPLUG HER PUMP IN STYLE POWER ADAPTER, A BLUE SPARK CAME FROM THE TRANSFORMER HOUSING AND THE POWER ADAPTER WAS HOT TO THE TOUCH WHEN SHE WENT TO REMOVE IT FROM THE WALL OUTLET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 174121 | PUMP IN STYLE (ITEM # UNK) | HGX | MEDELA, INC. | UNK/9207010 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |