FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE (ITEM # UNK)

MDR report key: 3843571 · Received March 25, 2014

Report

Report Number
1419937-2014-00295
Event Type
Malfunction
Date Received
March 25, 2014
Date of Event
March 10, 2014
Report Date
March 10, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER'S REPORT OF A SPARK COMING FROM THE HOUSING OF HER POWER ADAPTER IS A SAFETY ISSUE. IN F/U WITH THE CUSTOMER, THE CUSTOMER STATED THAT THERE WERE NO INJURIES THAT RESULTED AND THAT SHE DISCARDED THE POWER ADAPTER. AS A RESULT OF CAPA (B)(4) WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES, A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. ACTIVITIES RELATED TO THIS NOTIFICATION ARE ON-GOING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT WHEN SHE WENT TO UNPLUG HER PUMP IN STYLE POWER ADAPTER, A BLUE SPARK CAME FROM THE TRANSFORMER HOUSING AND THE POWER ADAPTER WAS HOT TO THE TOUCH WHEN SHE WENT TO REMOVE IT FROM THE WALL OUTLET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174121 PUMP IN STYLE (ITEM # UNK) HGX MEDELA, INC. UNK/9207010 UNK

Patients

Seq Age Sex Outcome Treatment
1