MDR report key: 3843538 · Received March 25, 2014

Report

Report Number
1419937-2014-00293
Event Type
Malfunction
Date Received
March 25, 2014
Date of Event
March 17, 2014
Report Date
March 17, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K301614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER WAS SENT A REPLACEMENT POWER SUPPLY. THE CUSTOMER REPORTED A BREACH IN THE TRANSFORMER HOUSING. SHE DID NOT REPORT OF ANY FIRE, SPARK OR INJURY. THE PRODUCT INVOLVED IN THE COMPLAINT WAS NOT RETURNED FOR EVALUATION/INVESTIGATION AT THIS TIME. THEREFORE, NO CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF THE EVENT. SHOULD ADD'L INFO OR THE ORIGINAL PRODUCT BE RECEIVED, RESULTING IN NEW, CHANGED, OR CORRECTED INFO, A FOLLOW-UP REPORT WILL BE FILED AT THAT TIME. THIS ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING., HANDLING AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STAKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SVC THE TRANSFORMER HOUSING OF HER PUMP IN STYLE BREAST PUMP COMING APART EXPOSING THE UNDERLYING ELECTRONICS WHICH IS A SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174186 PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE HGX MEDELA, INC. 57063/9207010 1212/REV M

Patients

Seq Age Sex Outcome Treatment
1