FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 3843389 · Received March 14, 2014

Report

Report Number
1720753-2014-02337
Event Type
Malfunction
Date Received
March 14, 2014
Date of Event
February 21, 2014
Report Date
March 14, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CONTACTS ON THE CONTROL PANEL PCB WERE CLEANED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED A GENERATOR ERROR MESSAGE, WHICH WILL LIKELY RESULT IN A SYSTEM LOCK UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152420 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1