FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3843284 · Received June 2, 2014

Report

Report Number
1416980-2014-17640
Event Type
Malfunction
Date Received
June 2, 2014
Date of Event
May 8, 2014
Report Date
May 8, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) HAS BEEN RETURNED AND A DEVICE ANALYSIS IS CURRENTLY UNDERWAY. UPON COMPLETION OF THE EVALUATION OR IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). AN EXTERNAL/INTERNAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. TEMPERATURE CONFIRMATION TESTING WAS PERFORMED AND THE RESULTS WERE FOUND TO BE WITHIN SPECIFICATIONS. THE DEVICE WAS DETERMINED TO MEET ELECTRICAL PERFORMANCE SPECIFICATION REQUIREMENTS PER RITE (RETURNED INSTRUMENT TEST EVALUATION) TESTING. THE DEVICE FAILED RITE FUNCTIONAL TESTING AS FLUID WAS TRANSFERRED ABOVE THE ALLOWABLE SPECIFICATION RANGE DURING VOLUMETRIC ACCURACY TESTING. TEST ARTICLE, PISTON FOAM, WAS INSTALLED AND THE DEVICE PASSED SUBSEQUENT VOLUMETRIC ACCURACY TEST. THE EVALUATION REVEALED THE CAUSE OF THE FAILURE TO BE DETERIORATED PISTON FOAM. THE PISTON FOAM WAS TO BE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED THERAPY MONITORED FLUID VOLUME TESTING. THERE WAS NO PATIENT INVOLVED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321221 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1