FDA Adverse Event
Malfunction
Summary report: N
VASOVIEW HEMOPRO 2
MDR report key: 3843239
·
Received March 14, 2014
Report
- Report Number
- 2242352-2014-00256
- Event Type
- Malfunction
- Date Received
- March 14, 2014
- Date of Event
- February 18, 2014
- Report Date
- February 18, 2014
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K101274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- PHYSICIAN ASSISTANT
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED SIGNS OF CLINICAL USAGE AND EVIDENCE OF BLOOD. A VISUAL INSPECTION DETERMINED THAT THE HEATING WIRE WAS DETACHED. ADDITIONAL EVALUATION IS REQUIRED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE REPORTED IN THE LOT HISTORY. (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO 2 OVERHEATED AND WOULD NOT TURN OFF. A REPLACEMENT DEVICE AND POWER SUPPLY WAS USED TO COMPLETE THE CASE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 152838 | VASOVIEW HEMOPRO 2 | ENDOSCOPIC VESSEL HARVESTING | GEI | MAQUET CARDIOVASCULAR, LLC | VH-4000 | 25090520 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR |